CI-PC36 dew point analyzer is an intelligent dew point analyzer with microprocessor as its core and ultra-thin alumina capacitive sensor as its measuring unit. It has the characteristics of high precision, good stability, comprehensive function and long service life.
Measure range |
Basic type : -60 - + 20℃ |
Standard type |
-80 - + 20℃ |
Extended type |
-100- + 20℃ |
Accuracy |
±3℃(-80- -60℃), ±2℃(-60 - + 20℃) |
Repeatability |
±0.5℃ |
Zero drift |
±1% FS/7d |
Range drift |
±1% FS/7d |
Response time : |
At the temperature of +20℃ and pressure of 1bar, the range reaches 63%[90%] |
-60 - -20℃ Td |
5s[10s] |
-20 - -60℃ Td |
45s[10min] |
Power supply |
85-264VA, 50/60Hz |
Environment temperature |
0-45℃ Environment Humidity: <90%RH |
Sample temperature |
0-50℃ |
Sample gas pressure |
Atmospheric conditions±10% (Gas outlet should be atmospheric conditions) |
External control signal |
1 set, control signal: 9-28VDC(optional) |
Analog output |
Standard confiquration set of 4-20MA, 0-20mA and 0-10V can be set by menu |
Communication |
RS485 (standard) / RS232 (optional) |
Alarm outout |
2 sets |
Relay contact point capacit |
24VDC, 0, 0.2A |
Inlet and outlet |
1 / 4 inch NPT internal thread, standard 6.5 hose ioin |
Instrument size |
W*H*D = 144m*144mm*251mm |
Immnstallation hole size of instrument |
L*W = 136mmx136mm |
USER EVALUATE
FAQ
The target market of our brand has been continuously developed over the years. For more information, please reach our team, we are more than happy to help.
1) O2/N2/H2 Flow rate: ____Nm3/h
2) O2/N2/H2 Purity: ____%
3) O2/N2/H2 Discharge pressure: ____Bar
4) Voltages and Frequency: _____V/PH/HZ
5) Application or use for which industry.
We will recommend the most suitable equipment for you according to your needs.